Skip to main content

PROJECT EXPERIENCE SECTION OF THE QSTI/QSTO

PROJECT EXPERIENCE SECTION OF THE QSTI APPLICATION - EXAMPLE  - For those persons who have taken the exam (or plan to) and are preparing their applications for submittal to the QSTI Review Committee, the project experience section may prove a little open-ended. What are “they” looking for? How much information should I include? Do “they” want a copy of a test report? First, definitely do not send in any test reports. The committee is not equipped to handle such confidential business information. What the committee does need is information about your experience and knowledge sufficient to demonstrate that you can conduct a test project successfully and produce reliable results. As a guideline for completing this element of the application, the QSTI Review Committee has provided two examples for your use. The components of these examples coincide with those outlined in the SES application form. Group II Application (example 1)  Project One  EPA Method 7  I led a project to measure NOx em

How Work Flame Ionization Detection (FID) in TVA-1000B?

Flame Ionization Detection (FID)

A Flame Ionization Detector (FID) measures organic compounds by utilizing a flame produced by the combustion of hydrogen and air. When hydrocarbons in the sample are introduced to the detection zone, ions are produced by the following reaction:
RH + O → RHO+ + e– → H2O + CO2
where
R = carbon compound
A collector electrode with a polarizing voltage is also located within the detector chamber, and the ions produced by this reaction are attracted to it. As the ions migrate towards the collector,a current is produced which is directly proportional to the concentration of hydrocarbons introduced to the flame. This current is then amplified and sent to a microprocessor and/or analog readout device.
The FID has a wide dynamic range. The effective dynamic range can be further expanded by use of a dilutor kit which reduces very high volatile organic compounds (VOC) concentrations to within the dynamic range (or even linear range) of the analyzer. The dilutor kit can also be used to enrich oxygen deficient samples by adding ambient air that is rich in oxygen (20.9% usually). Low oxygen can affect the characteristics of the hydrogen flame, causing readings to be artificially elevated and possibly extinguishing the flame. As a general rule of thumb, greater than 16% oxygen is required to support the flame. If underground gases or samples in gas bags are to be measured by an FID, it is advised that the dilutor be used to combat the problem.

Benefits of Flame Ionization Detection

♦ Wide dynamic and linear range
♦ High sensitivity to hydrocarbon vapors (including methane)
♦ Very stable and repeatable response
♦ Virtually unaffected by ambient levels of CO, CO2, and water vapor




Comments

Post a Comment

Popular posts from this blog

Sources of Air Pollution are four mobile sources, stationary sources,point sources andarea sources

 Sources of Air Pollution In discussions of air pollution sources, four terms are routinely used:  Mobile sources  Stationary sources  Point sources  Area sources  Mobile sources Mobile sources are generally related to transportation. Mobile sources are not included within the general parameters for fugitive emissions and, therefore, will be given no further consideration in this course.  Stationary sources Stationary sources define emission sources that are fixed and include all forms and types of manufacturing, chemical and petroleum processing, power generation, and small business facilities. Point sources A stationary point source refers to any place within a facility from which emissions are possible. For a given facility, a stack may be one point source. An emission leak within a process is also considered a point source. An automobile or a truck would represent a point source or mobile source. Area sources Area source refers to any one of several sources contributing to the air

PROJECT EXPERIENCE SECTION OF THE QSTI/QSTO

PROJECT EXPERIENCE SECTION OF THE QSTI APPLICATION - EXAMPLE  - For those persons who have taken the exam (or plan to) and are preparing their applications for submittal to the QSTI Review Committee, the project experience section may prove a little open-ended. What are “they” looking for? How much information should I include? Do “they” want a copy of a test report? First, definitely do not send in any test reports. The committee is not equipped to handle such confidential business information. What the committee does need is information about your experience and knowledge sufficient to demonstrate that you can conduct a test project successfully and produce reliable results. As a guideline for completing this element of the application, the QSTI Review Committee has provided two examples for your use. The components of these examples coincide with those outlined in the SES application form. Group II Application (example 1)  Project One  EPA Method 7  I led a project to measure NOx em

Dual Detectors FID and PID

Dual Detectors The benefits of each individual detector are very clear: both the FID and the PID have their  advantages and disadvantages. However, with either detector alone, the number of organic  and inorganic vapors that one can detect is limited by the measurement capabilities of that  detector. With the TVA-1000B, users can obtain complete information about more organic and inorganic  vapors more quickly and easily than with single detector technology alone. Since both detectors may be displayed and logged simultaneously, the relative response of the  two detectors may give some clues about the identity of the compound being measured. For  instance, the PID does not respond to methane at all, but the FID responds very well. A high  FID reading with virtually no PID response might indicate the presence of methane. Consequently,  PIDs respond very well to some inorganic gases that FIDs cannot detect. A high PID  reading with no FID reading might suggest the presence of an i