PROJECT EXPERIENCE SECTION OF THE QSTI APPLICATION - EXAMPLE  - For those persons who have taken the exam (or plan to) and are preparing their applications for submittal to the QSTI Review Committee, the project experience section may prove a little open-ended. What are “they” looking for? How much information should I include? Do “they” want a copy of a test report? First, definitely do not send in any test reports. The committee is not equipped to handle such confidential business information. What the committee does need is information about your experience and knowledge sufficient to demonstrate that you can conduct a test project successfully and produce reliable results. As a guideline for completing this element of the application, the QSTI Review Committee has provided two examples for your use. The components of these examples coincide with those outlined in the SES application form. Group II Application (example 1)  Project One  EPA Method 7  I led a project to measure NOx em

What is Photo Ionization Detector (PID)?

Photoionization Detector (PID)


A Photoionization Detector (PID) consists of an ultraviolet (UV) lamp of a specific energy and an ionization chamber. Compounds passing through the chamber are excited by photons of UV energy and ionized according to the following equation:
R + hυ → R+ + e–
where
R = most organic/inorganic compounds
These ions are attracted to a collecting electrode, producing a current proportional to the concentration of the compound.
Whether or not a compound can be detected by a PID depends upon the energy required to remove an electron from the compound (its ionization potential). If the lamp energy is greater than the compound’s ionization potential, the PID will detect it. The standard lamp in the TVA-1000B is 10.6 eV. Other lamps (9.6 and 11.8 eV) are also available. The 11.8 eV lamp permits detection of many compounds not ionized by the standard lamp. The lower energy (10 eV) lamps, however, allow more selectivity by not responding to undesired compounds with a higher ionization potential.

NOTE: Refer to MI 611-183 for information and guidance on proper use of the 11.8 eV lamp.

Because of its smaller dynamic range (0-2000 ppm), the PID is not the detector of choice for measuring high concentrations of vapors. A PID is also more susceptible to interference from water vapor than a FID. However, as a PID does not require hydrogen or oxygen, it is the detector of choice when fuel is limited or unavailable, or when ambient oxygen concentrationsare low. The PID is also very sensitive to aromatic and chlorinated compounds, and can even measure some inorganic compounds that the FID does not detect at all (ammonia, carbon disulfide, carbon tetrachloride, chloroform, ethylamine, formaldehyde, and hydrogensulfide, to name a few).

Benefits of Photoionization Detection

♦ High sensitivity to aromatics, unsaturated hydrocarbons and chlorinated hydrocarbons
♦ Ability to measure some inorganic gases
♦ Very simple operation
♦ No support gases required
♦ Non-destructive detector allows sample to be recovered



Comments

Popular posts from this blog

Sources of Air Pollution are four mobile sources, stationary sources,point sources andarea sources

 Sources of Air Pollution In discussions of air pollution sources, four terms are routinely used:  Mobile sources  Stationary sources  Point sources  Area sources  Mobile sources Mobile sources are generally related to transportation. Mobile sources are not included within the general parameters for fugitive emissions and, therefore, will be given no further consideration in this course.  Stationary sources Stationary sources define emission sources that are fixed and include all forms and types of manufacturing, chemical and petroleum processing, power generation, and small business facilities. Point sources A stationary point source refers to any place within a facility from which emissions are possible. For a given facility, a stack may be one point source. An emission leak within a process is also considered a point source. An automobile or a truck would represent a point source or mobile source. Area sources Area source refers to any one of several sources contributing to the air

PROJECT EXPERIENCE SECTION OF THE QSTI/QSTO

PROJECT EXPERIENCE SECTION OF THE QSTI APPLICATION - EXAMPLE  - For those persons who have taken the exam (or plan to) and are preparing their applications for submittal to the QSTI Review Committee, the project experience section may prove a little open-ended. What are “they” looking for? How much information should I include? Do “they” want a copy of a test report? First, definitely do not send in any test reports. The committee is not equipped to handle such confidential business information. What the committee does need is information about your experience and knowledge sufficient to demonstrate that you can conduct a test project successfully and produce reliable results. As a guideline for completing this element of the application, the QSTI Review Committee has provided two examples for your use. The components of these examples coincide with those outlined in the SES application form. Group II Application (example 1)  Project One  EPA Method 7  I led a project to measure NOx em

Dual Detectors FID and PID

Dual Detectors The benefits of each individual detector are very clear: both the FID and the PID have their  advantages and disadvantages. However, with either detector alone, the number of organic  and inorganic vapors that one can detect is limited by the measurement capabilities of that  detector. With the TVA-1000B, users can obtain complete information about more organic and inorganic  vapors more quickly and easily than with single detector technology alone. Since both detectors may be displayed and logged simultaneously, the relative response of the  two detectors may give some clues about the identity of the compound being measured. For  instance, the PID does not respond to methane at all, but the FID responds very well. A high  FID reading with virtually no PID response might indicate the presence of methane. Consequently,  PIDs respond very well to some inorganic gases that FIDs cannot detect. A high PID  reading with no FID reading might suggest the presence of an i